Logic Class II, Type A2 Biosafety Cabinets (BSCs) provide personnel, product and environmental protection from hazardous particulates, including agents requiring BSL 1, 2, or 3 containment.These cabinets are also suitable for handling antineoplastic drugs, genetic material, carcinogens, allergens and other substances that generate hazardous airborne particulates.
Cell Logic Class II, Type A2 BSCs are designed to meet the specific needs of cell research.Some models feature a modified sash to accommodate a microscope, and others offer a temperature-controlled work area on the work surface to maintain cell viability.
Both Logic and Cell Logic Class II, Type A2 BSCs utilize the same airflow pattern to ensure personnel and product protection.During operation, room air is drawn into the inlet grille at the work access opening and passes through a supply HEPA filter.Approximately 70% of the HEPA-filtered air is recirculated through the cabinet, while 30% passes through an exhaust HEPA filter and is discharged.
SEE OUR GREEN INITIATIVES
Features (All Models)
- Long-life DC Electronically Commutated Motor (ECM)
- Constant Airflow Profile™ (CAP) airflow monitoring system
- BalanceFlow™ plenum provides energy-efficiency and ultra-quiet operation*
- Easy clean top cover that promotes even exhaust airflow from the top cover
- 必威平台官网首页Mag-Zone™ area on interior right side manages storage challenges with Labconco Magnetic Mag-Zone accessories
- Intrinsically-safe negative pressure design
- Fully scannable 99.99+% supply and exhaust HEPA filters.99.999% efficient ULPA filters or H14 HEPA filters also available
- Logic OS and 5" interior-mounted, line-of-sight touchscreen information center
- Smart-Start™ System allows user to start up and shut down cabinet functions by raising and lowering the sash
- Night-Smart™ System for idling the blower when the sash is fully closed
- Full operational maintenance and certification logs for detailed record keeping
- Status line for alarm conditions and alerts to warn when filter life diminishes to 20% and 10% before alarming at 0%
- Interval/elapsed timer for experiment monitoring
- UV light timer (on models with UV)
- Password protection for cabinet settings
- 8 Languages
- Bright, glare-free LED lighting up to 180 foot-candles (1937 lux).Located outside the contaminated work area
- Curved, stainless steel inlet grille with waterfall design and Reserve-Air™ Secondary Airflow Slots
- Air-Wave™ Entry System at access opening to enhance containment
- Contain-Air™ Negative Pressure Channel at top of sash to enhance containment
- Two electrical outlets.Flush mounted stainless steel cover with dampened hinges
- 115V GFCI, double receptacles, maximum amperage draw 5A
- 230V, single receptacles, maximum amperage draw 3A
- 10° angled, fully closing sash.1/4" tempered safety glass with counterbalanced, anti-racking, easy-to-lift mechanism
- Unitized, pressure tested stainless steel interior.Optic White epoxy-coated steel exterior
- Removable towel catch located under work surface
- 8", 10", or 12" (20.3, 25.4, or 30.5 cm) working sash height
- 22.5" (57 cm) max.sash opening & 27.0" (69 cm) viewing height
- Removable, seamless type 304 stainless steel, dished work surface with lift out knobs
- Nominal inflow velocity of 105 fpm (0.53 m/sec)
- Nominal downflow velocity of 55 fpm (0.28 m/sec).6' x 12" sash models nominal downflow velocity of 50 fpm (0.25 m/s)
- Approximately 70% air recirculation
- Noise level <60 dB (4ft 10" model)
- 10' (3 m) power cord with plug
- Six year warranty
Options
- 3', 4', 5' and 6' widths
- International electrical configurations
- Unassembled, NSF-approved, epoxy-coated steel, non-welded telescoping base stand with leveling feet
- Accessory Package: UV lamp (254 nm), right-side service fixture with ball-type valve and right and left-side Vacu-Pass™ Cord & Cable PortalCell Logic+ packages specifically designed for use with microscopes and temperature-sensitive samples
- Ventus™ II Canopy Connection
- Cell Logic packages specifically designed for use with microscopes and temperature-sensitive samples
Required Accessories (Not Included)
- Supporting base if non-welded stand option is not selected
Compliance
- NSF/ANSI Standard 49 listed
- NSF/ANSI 49 compliant (Cell Logic models)
- ISO Class 5 (Grade A) per ISO 14644-1
- UL Standard 61010-1
- CAN/CSA 22.2 No.61010-1
- CE/UKCA Mark (230V models)
- ADA-compliant